Littlebrook specialize in the calibration of air temperature measurement instrumentation . Thermometry used with Stack/Emissions pitot static tubes ideally should ideally be calibrated as close as possible to their conditions of use.

Littlebrook have developed Accredited Temperature Calibration Procedures using over fluidized power baths to simulated temperatures in a “Dirty Stack”.

Measured Quantity
Instrument or Gauge

Range Best Measurement
Capability Expressed
as an Expanded
Uncertainty (k=2)
TEMPERATURE    
Triple point of Mercury
Triple point of water
Melting point of Gallium
Freezing point of Indium
Freezing point of Tin
Freezing point of Zinc
-38.8344 °C
0.01 °C
29.7646 °C
156.5985 °C
231.928 °C
419.527 °C
0.003 °C
0.002 °C
0.003 °C
0.004 °C
0.005 °C
0.006 °C
Calibration in Air Chamber

0 °C to 90 °C

0.4 °C to 0.6 °C

Calibration in fluidised powder bath 50 °C to 300 °C
0.5 °C
Platinum Resistance Thermometers and Electronic Thermometers with sensors
-90 °C to -40 °C
-40 °C to 0 °C
0 °C to 100 °C
100 °C to 155 °C
155 °C to 320 °C
320 °C to 650 °C
0.05 °C
0.025 to 0.021 °C
0.019 °C
0.019 to 0.025 °C
0.09 °C
0.14 °C
Thermocouples, base metal -90 °C to 650 °C 0.5 °C
Metal block calibrators -90 °C to 155 °C
155 °C to 320 °C
320 °C to 650 °C
0.06 °C
0.10 °C
0.14 °C
HUMIDITY    
Relative Humidity Instruments 10 % rh
35 % rh
50 % rh
65 % rh
75.3 % rh
80 % rh
1.4 % rh
1.9 % rh
2.0 % rh
2.2 % rh
2.2 % rh
2.3 % rh
For full Accreditation details see www.ukas.org